Display Features
The monitor displays tpO
2
/CO
2
values and trends as follows:
In the tpO
2
/CO
2
parameter box on the Main screen
In the trend display (see chapter
On the oxycardiorespirogram (see page 12-9)
When two tpO
2
/CO
2
pods are connected, each displays a parameter box on the main
screen. The labels of the second pod are marked by an asterisk (*), as shown below:
tpO
2
/CO
2
Setup
High sensor temperatures on the skin’s surface offer the greatest degree of accuracy
and can quickly detect a change in the patient’s condition. However, the heat from the
sensor may cause severe burns, especially in neonates. On the tpO
2
/CO
2
setup menu,
you can restrict the application time of the sensor and establish a temperature set point
(the target temperature for heating the sensor during calibration).
The tpO
2
/CO
2
pod's settings for altitude and barometric pressure influence the
measurement. The tpO
2
/CO
2
pod contains an internal pressure sensor that measures
current barometric pressure, which changes with altitude and weather conditions. The
monitor allows you either to rely on the barometric pressure values generated
automatically by the pod or manually enter pressure values obtained with an external
barometer. Readings are displayed on the tpO
2
/CO
2
setup menu.
See “Precautions” on page 18-3 for other important safety information.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...