![Dräger Infinity Delta Instructions For Use Manual Download Page 310](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779310.webp)
17-20
D
ELTA
/D
ELTA
XL/K
APPA
VF8
17 P
ULSE
O
XIMETRY
(S
P
O2)
®
Standalone Pulse Oximeter
Overview
The can be used as a second source of SpO
2
monitoring with monitors that
are compatible with Dräger Oxisure via MultiMed. It is a small, battery operated pulse
oximeter that connects to the monitor's X8 connector via an RS232 cable.
When monitoring SpO
2
via the the parameter box is labeled SpO2*.
SpO
2
*alarms are set from the Alarm Limits menu (see chapter 5). SpO
2
* does not
display a waveform and is not supported for OCRG.
NOTE:
You cannot use the remote keypad and/or the Masimo SET or Nellcor Oximax
pods with the simultaneously.
pulse oximeter is not supported on monitors with Masimo or Nellcor
logo (see pages 17-4 and 17-5).
SpO
2
connector
Kappa rear
Delta/Delta XL right side
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...