S
TATUS
M
ESSAGES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
20-9
etCO
2
Cal. Failed
Calibration failed.
• Retry calibration.
• Return the pod to hospital’s technical
personnel.
etCO
2
pod
Disconnected
The sidestream pod is unplugged
from the monitor.
• Check cable and connection.
• Replace the cable, if necessary.
etCO
2
pod
Incompatible
Corrupt EEPROM. The wrong
software or hardware version.
• Try a new pod.
• Consult the hospital’s technical
personnel.
Disconnect
Duplicate etCO
2
pods
Multiple pods are connected.
Disconnect unwanted duplicate pod.
Service etCO
2
pod
Pod must be serviced if pump hrs. >=
20,000 hrs.
Contact DrägerService or replace pod.
CO2: H/W
Problem
Microstream pod hardware
malfunction.
Disconnect the Microstream pod, then
reconnect it. If the message persists
return the pod to the hospital’s
technical personnel.
Message
Condition
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...