Other ECG Setup functions
HR Alarm
Accesses Alarm Limits table
See chapter 5, Alarms, for more
information about setting and
displaying alarm limits.
• Click on
HR Alarm
to open the Alarm Limits
table with HR-associated alarms prioritized.
WARNING
: When the HR alarm and the ARR
monitoring functions are set to
OFF
, the monitor
cannot generate ASY/VF alarms. To restore this
ability, set ASY/VF alarms to
Always ON
(see
page 2-14).
HR Source
Selects the source for the heart
rate (especially useful during
electrosurgery when artifact
interferes with ECG heart rate
detection).
NOTES
: When the monitor is
part of a network, the rest of the
system continues to display the
HR label in the ECG parameter
box, regardless of the source.
For example, even if you select
SpO
2
as the HR source at the
monitor, the Infinity
CentralStation displays HR in
the ECG parameter box.
• Click on
ECG
to derive heart rate from ECG
signal.
• Click on
ART
to derive heart rate from Arterial
Pressure signal. HR parameter box label
changes to APR and displays values in red. If
the monitor cannot detect a signal, it defaults
to ECG for the heart rate.
• Click on
SpO2
to derive heart rate from pulse
oximetry signal. HR parameter box label
changes to PLS and displays values in white.
The pulse visual blip and audio tone show no
change indicating SpO
2
saturation values.
• Click on
AUTO
to derive heart rate from ECG
signal or other available signals. If an ECG
signal is unavailable, the monitor switches to
ART, then SpO
2
. The pulse blip and audio
tone are derived from the same parameter as
the selected HR source.
QRS/ARR
Select
Facilitates accurate detection of
HR and ARR calls by allowing
you to select single- or dual-
channel processing for
maximum signal clarity. This
setting affects how the monitor
responds to artifact.
NOTES
:
• How the monitor responds to
artifact depends on whether
ECG monitoring is configured
for single- or dual-channel
processing (see page 8-18).
• Regardless of your
configuration of this setting,
the monitor resumes QRS
processing but does not
initiate a relearn when an
artifact clears.
• The
QRS/ARR Select
function
is ghosted in neonatal mode.
• Click on
ECG 1
to determine heart rate and
arrhythmias based on the single best lead.
• Click on
ECG1&2
to determine heart rate and
arrhythmias based on the two best leads.
Menu
Selection
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...