![Dräger Infinity Delta Instructions For Use Manual Download Page 556](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779556.webp)
B-30
D
ELTA
/D
ELTA
XL/K
APPA
VF8
B T
ECHNICAL
D
ATA
tpO
2
/CO
2
Pod
FiO
2
Sensors
Physical Attributes
Size (H x W x D):
52 x 132 x 153.4 mm (2.05 x 5.2 x 6.04 in)
Weight:
454 g (1 lb)
Connections:
Connector for tpO2/tpCO2 solid-state electrode
Connector for pod-monitor interface cable
Interface Cable:
1 m
Degree of protection against ingress of
water:
per IEC 60601-2-27 clause 44.3
Electrical Specifications
Mode of operation:
Continuous
Power requirements:
DC powered directly from the monitor
Protection against electric shock:
Type CF
Environmental Requirements
Temperature range:
Operating:15 to 40 °C (55 to 104 °F)
Storage: –20 to 50 °C (–-4 to 122 °F)
Relative humidity:
Operating:20 to 95 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atmospheric pressure:
Operating:550 to 775 mmHg (73 to 103 kPa)
Storage: 375 to 795 mmHg (50 to 106 kPa)
Physical Attributes
Size (H x D):
40 x 30 mm (1.59 x 1.2 in)
Weight:
35 g (<1.2 oz)
Connections:
NeoMed Pod, interface cable
Mounting:
16 mm thread x 1 mm pitch
Sensor type:
Galvanic fuel cell (partial pressure)
Useful life:
Approximately 1 year
NOTE
: This sensor contains lead and caustic material. Dispose of or recycle properly and in
accordance with local regulations.
Environmental Requirements
Temperature range:
Operating:10 to 40 °C (50 to 104 °F)
Storage: –10 to 50 °C (14 to 122 °F)
Relative humidity:
Operating:20 to 90 %, non-condensing
Storage: 10 to 95 % (with packaging)
Atm. pressure:
600 to 900 mmHg (80 to 120 kPa)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...