ECG P
RECAUTIONS
VF8
D
ELTA
/D
ELTA
XL/K
APPA
8-7
To enable the ESU Filter for ESU Block or
MultiMed Plus OR Cable
1. Click on the
HR
parameter box on the main
screen.
2. Click on
ECG Options
.
3. Click on
Filter
.
4. Scroll to
ESU
and click.
5. Plug the ESU 5 block into the MultiMed 5 pod.
When using a MultiMed 6 pod, leave the V+ lead
connection open.
6. Plug in the applicable ECG lead set (see appendix C, “Approved Options and
Infusion or Roller Bypass Pumps
Infusion or roller bypass pumps may cause artifact in the monitor's ECG signals. Such
interference may cause the monitor to display pacemaker spikes even though the ECG
waveform appears normal. To determine if the pump is the source of the electrical
disturbance, turn it off, if possible. If the artifact disappears, the pump is the probable
cause. To reduce such artifact, choose the lead with the best signal for monitoring or
replace the electrodes. Rerouting invasive pressure tubing away from the infusion
pump tubing may also improve the signals.
Line Isolation Devices
To reduce the effect of line isolation devices, which may cause temporary
disturbances (transients) in the ECG signal, take the following precautions:
Choose the lead with the best signal for monitoring.
Check the electrodes and replace them, if necessary.
NOTE:
The MultiMed Plus OR cable has an integrated
ESU filter. Therefore, no ESU block has to be connected.
WARNING: During electrosurgery, Dräger recommends using
an ESU block or an MultiMed Plus OR cable with Dräger ECG
lead sets. They help protect the patient from burns caused by
ESU-induced current flowing through the leads.
ESU 5 Block
MultiMed 5 Pod
5 lead set
ESU 6 Block
6 lead set
MultiMed 6 Pod
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...