MONITORING
Central Venous pressure
Check for accurate CVP signal and values before performing TD
measurement and calibration of Pulse Contour parameters.
Amount of Bolus
15 cc is normally adequate for most adult patients but can be changed as
necessary.
Temperature of Injectate
The signal is better with iced temperature injectate or cold temperature
injectate (refrigerated) or at least 10
o
C below blood temperature as opposed
to warmer/room temperature injectate. The reason for this is that the bolus has
to traverse a longer path (especially femoral catheter) before being detected at
the blood temperature thermistor end (as opposed to regular RH
Thermodilution CO where the path is much shorter). If warmer temperature
injectate is used, the difference in temperature will be very small, making it
hard to calculate a good CO measurement.
In case there is no pulmonary edema, room temperature injectate can be used.
Usually, warmer temperature injectate could be used if EVLWI <10.
If individual measurements are not in good agreement and the TD peak is low,
< 0.15 °C (59 °F), it is good to use iced injectate or at least injectate coming
out of the refrigerator.
Injectate temperature generally must not be warmer than 24 °C (75 °F). If
injectate temperature is above 31°C (88°F) or fluctuates and is invalid, no
measurement will be possible. An error message
Check Injectate Probe
will
be displayed and the p-IT values will display ***. Before starting a
measurement, make sure the thermistor is not exposed to an external heat
source like a warming blanket, etc. Flushing with a cold saline before the first
measurements helps to overcome these issues.
Injection time
The injection time must be within 7 sec. The injection should be done as
quickly as possible via the distal CV lumen.
A shorter injection makes a better signal. Also, the injection should be done as
smoothly as possible.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...