20-2
D
ELTA
/D
ELTA
XL/K
APPA
VF8
20 M
ICROSTREAM
®
ET
CO2 M
ONITORING
Overview
The Infinity
®
etCO
2
Microstream
®
SmartPod
®
, when connected to a monitor, allows
the monitor to measure and display sidestream etCO
2
(end-tidal CO
2
), iCO
2
(inspired
CO
2
) and RRc (Respiration Rate). The Microstream pod cannot be used
simultaneously with either an etCO
2
pod or an etCO
2
/respiratory mechanics pod.
The Microstream pod is intended for use with adult, pediatric, and neonatal patients
with appropriate accessories (see page C-24). The pod uses a fixed sample flow rate of
50 mL/min. and requires a proprietary sample line (accessory) and connects to the
monitor via PodCom.
Precautions
Refer to the general precautions section for etCO
2
on page 19-2.
CAUTION:
To avoid blocking the sampling tube, do not attempt to sterilize the
Microstream pod by immersing it in liquids.
Do not obstruct the exhaust port on the Microstream pod because it prevents
air flow.
NOTE:
The pod and associated accessories that have patient contact are free of latex.
The pod automatically compensates for atmospheric pressure within the defined
operating ranges listed in appendix B.
The Infinity etCO
2
Microstream SmartPod meets its accuracy specifications
within 1.1 minutes after the monitor is powered on.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...