15 C
ARDIAC
O
UTPUT
(C.O.)
Catheters (Comp. Constant)
The monitor compensates for discrepancies in catheters used to measure C.O. The
catheter compensation factor is listed as Comp. constant on the C.O. setup menu.
If you use an Edwards/Baxter, BD/Ohmeda, or Arrow catheter, the computation
constant is automatically chosen for you. You can, however, enter a different value
(whenever, for instance, you change either injectate volume or temperature).
Choosing a catheter type determines the available choices under
Catheter Size
and
Injectate Volume
. The following tables list the computation constants for Edwards/
Baxter, BD/Ohmeda, and Arrow catheters.
WARNING: An incorrect computation constant may yield
incorrect C.O. measurements and put the patient at risk.
Confirm that the manually entered computation constant is
correct for the catheter you are using.
Edwards/Baxter
Injectate Temperature (IT)
Sensor connected
IT Sensor
disconnected
Catheter
Size
Injectate
Volume
IT = -5° to +16°
IT = 16° to 27°C
IT = 0°C
7F
10 cc
0.561
0.608
0.542
7F
5 cc
0.259
0.301
0.247
7.5F
10 cc
0.574
0.595
0.564
7.5F
5 cc
0.287
0.298
0.257
5F
5 cc
0.285
0.307
0.270
BD/Ohmeda
Injectate Temperature (IT)
Sensor connected
IT Sensor
disconnected
Catheter
Size
Injectate
Volume
IT = -5 °C to +16 °C
IT = 16 °C to 27 °C
IT = 0 °C
7.5F
10 cc
0.579
0.628
0.566
7.5F
5 cc
0.281
0.309
0.270
7.5F
3 cc
0.160
0.181
0.151
7F
10 cc
0.579
0.628
0.566
7F
5 cc
0.281
0.309
0.270
7F
3 cc
0.160
0.181
0.151
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...