The Menu Setup submenu
Menu Setup
Determines amount of time menus and
screens remain displayed
NOTE
: This setting determines menu
display time in
Remote View
also.
1) Click on
Menu Setup
.
2) Click on
Menu Time Limit
.
3) Click on one of the following:
•
ON
– Active menus and screens
are displayed for a limited time
only (approximately 5 minutes).
•
OFF
– Menus and screens
remain displayed until you
cancel them or select another
display.
The Save/Restore and Rename Setups submenus
These menus allow the unit manager to save and restore user-defined and default setups. Open
the Unit Manager menu (page 2-14), click on
Save/Restore
or
Rename Setups
, then follow the
procedures outlined on page 2-10 to execute the indicated functions.
The TPO
2
/CO
2
submenu
This menu configures the monitor for transcutaneous blood gas monitoring. Open the Unit
Manager menu (page 2-14), click on
TPO2/CO2
, select and execute functions as described in this
table. For more about tpO
2
/CO
2
functions, see page 18-10.
Site Timer
Control
Restricts or permits access to the site
timer (see page 18-11).
•
Nurse
– Clinical personnel can
set the site timer in the
tpO
2
/CO
2
menu.
•
Manager
– Access to the site
timer is restricted by the
password-protected Unit
Manager menu.
Site Timer
Limits the use of the tpO
2
/CO
2
sensor to a
specified period of time. When the time
has expired, an alarm sounds and all
network devices are advised with a
message.
•
0.5
-
8.0
hr
in 0.5 hr increments
• OFF
Timer OFF
Control
Enables/disables the control of the site
timer (see page 18-11).
•
ON
– Only the unit manager can
disable the site timer.
•
OFF
– The clinical personnel
can disable the site timer in the
tpO
2
/CO
2
menu.
Auto Heater
Shutdown
Refer to your hospital’s policy for this
feature.
NOTE
:
The sensor heater is disabled in standby
mode to conserve power and prolong the
life of the sensor.
•
ON
– Disables the sensor
heater when site timer expires.
•
OFF
– Continues tpO
2
/CO
2
monitoring when the site timer
expires.
The Unit Manager Menu
Menu Item
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...