When both batteries run low, the monitor sounds an alarm, and a status message
appears in the network message area. If both batteries are depleted, the monitor turns
off automatically.
The external battery fits into a compartment on the
monitor’s left side. When depleted or removed, replace
it immediately or connect the monitor to a power
supply. The battery is being charged (as indicated by the
battery charger LED on the front panel) whenever the
monitor is connected to AC power.
The internal battery is charged first, then the external
battery. The table below illustrates the function of the
battery charge bar graph at the top of the screen:
WARNING: Worn out or defective batteries can significantly
reduce these times.
CAUTION:
The battery charger display is only accurate if the batteries are in normal
working condition.
NOTE:
When AC power is disconnected, the battery charge display takes up to 15
seconds to reflect the internal battery's actual capacity and up to 60 seconds to
reflect the external battery's actual capacity.
The battery indicator is not displayed if the monitor is not connected to the
Infinity network and the setting
Network Mode
is set to
Direct Net Mode
and
the setting
Network Control
is set to
ON
. To display the battery indicator under
these circumstances, the setting
Network Control
must be set to
OFF
.
Battery Charge Display
Display
Charge Left
Action
The battery is fully charged
Not applicable
The battery is half full
Connect the monitor to an IDS or AC
adapter.
The external battery is very low
(< 25 %)
Replace with a fully charged external
battery.
The external battery is deplete.
1
Replace with a fully charged external
battery.
Insert
battery here
Battery
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...