9-4
D
ELTA
/D
ELTA
XL/K
APPA
VF8
9 A
RRHYTHMIA
M
ONITORING
Automatic Learning and Relearning
After you connect ECG cables to the patient, the monitor begins to learn a reference
template whenever you execute any of the following tasks:
Turn on the monitor
Exit standby mode
Click on
ARR Monitoring
or
QRS/ARR Select
Change the top-channel lead (ECG 1)
,
or change the ECG2 channel lead
during ECG1&2 processing
Change a 12-lead configuration
SVT
Supraventricular Tachycardia: N or more consecutive normal beats, with a beat-to-beat
rate greater than or equal to the SVT setting
CPT
Ventricular Couplet: Sequence of beats with the pattern: normal, PVC, PVC, normal
BGM
Ventricular bigeminy: Sequence of beats with the pattern: normal, PVC, normal, PVC,
normal
TACH
Sinus Tachycardia: N or more consecutive normal beats, with a beat-to-beat rate
≥
TACH
rate setting
2,3
BRDY
Sinus bradycardia: 8 or more consecutive normal beats, with an average rate
≤
sinus
bradycardia rate setting
2
NOTES
:
• When in neonatal mode, bradycardia is a low heart rate alarm
.
• Brady Alarm (neonate only), which is a high alarm, can be configured independently of
the low HR alarm, which is a medium alarm.
PAUS
Pause: Sequence of two beats classified as normal or PVC, with interval
≥
pause rate
value in seconds (±100ms)
ARTF
Artifact
:
More than 50 % of beats in the last minute classified as questionable
1
Certain ventricular tachycardias have sinusoidal waveforms closely resembling those of
ventricular fibrillation. Because of the similarity of these waveforms, the monitor may classify
such types of ventricular tachycardia as ventricular fibrillation, the more serious of the two
conditions.
2
“N” is the event count set in the Arrhythmia setup table’s count column.
3
A PVC or other abnormal beat breaks the analysis sequence and restarts analysis.
Label
Event and Beat Classification
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...