Measurement accuracy, Neonatal mode (1, 2):
Saturation (%SpO2):
0 to 60% not specified
60 to 80% sensor-specific as follows:
Nellcor :
OxiMAX MAX-N
±3
Nellcor :
SC-PR, SC-NEO, OxiCliq N, D-YS, OXI-A/N
not specified
70 to 100% sensor-specific as follows:
Nellcor:
Nellcor :
OxiMAX MAX-N, SC-PR, SC-NEO
±2
Nellcor :
OxiCliq N
±3.5
Nellcor :
D-YS, OXI-A/N
±4
Pulse Rate(3):
±3 beats/min or ±3% (whichever is greater)
NOTES:
1) Since pulse oximeter measurements are statistically distributed, only about two-thirds of those
measurements can be expected to fall within ±1
A
rms
of the value measured by a co-oximeter.
2) The Infinity® Nellcor Oximax® SpO2 SmartPod™ pulse oximeter with Adult sensors has been
validated in human blood studies on healthy adult volunteers in induced hypoxia studies in the
range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This variation
equals ±1
A
rms
of the value measured by a co-oximeter.
3) The pulse rate accuracy has been validated on healthy adult volunteers during induced hypoxia
studies in the range of 70-100% SpO2 against a laboratory co-oximeter and ECG monitor. This
variation equals ±1
A
rms
of the pulse rate value measured by the ECG monitor.
Nominal wavelength:
Red:
660 nm
IR:
910 nm
Parameter display:
tpO2, tpCO2 parameter values; plus selectable on oxy-
cardirespirogram
Sensing element:
tpO2/tpCO2 solid-state, dual-thermistor electrode
Measurement range:
tpO2:
0 to 800 mmHg
tpCO2: 0 to 200 mmHg
Resolution:
1 mmHg
Reporting interval:
once every 1.2 s
Accuracy:
tpO2:
40 to100 mmHg ±1 mmHg
<40 or >100 mmHg ±2 % or ±5 mmHg,
whichever is greater
tpCO2: 30 to 80 mmHg: ±1 mmHg
<30 or >80 mmHg ±3 mmHg
Drift:
O2:
±5 % over recommended calibration interval
CO2:
±10 % over recommended calibration interval
Non-linearity:
±5 mmHg
Hysteresis:
±5 mmHg
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...