![Dräger Infinity Delta Instructions For Use Manual Download Page 178](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779178.webp)
10-2
D
ELTA
/D
ELTA
XL/K
APPA
VF8
10 ST M
ONITORING
Overview
ST segment deviation is defined as the displacement (in mm) above or below the
isoelectric level. The measurement of deviation compares the isoelectric point to the
ST measurement point.
The isoelectric point defines the point of zero voltage (no electrical activity, 0 mm)
with a default position of 28 ms before the onset of the QRS complex on the
horizontal (time) axis. The
ST point
occurs in the ST segment between the QRS offset
(J point) and the T-wave, at a default position of 80 ms after the QRS offset. The
following figure illustrates a typical QRS complex.
The ST analysis feature examines QRS complexes classified as “normal” beats from
up to twelve selected ECG leads. The monitor learns each ST lead, combining the
measurements and features of normal beats into a composite (or average) QRS
complex. It derives the ST segment deviation from this average.
When ST monitoring is enabled, current ST values are trended and can be reviewed on
the trend display.
1
Isoelectric point
(Default = 28 ms before QRS onset)
5
ST deviation
2
QRS onset
6
ST measurement point
(Default = 80 ms after QRS offset)
3
Fiducial point
7
Isoelectric level
4
QRS offset
8
ST level
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...