![Dräger Infinity Delta Instructions For Use Manual Download Page 280](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779280.webp)
16-8
D
ELTA
/D
ELTA
XL/K
APPA
VF8
16 C
ALCULATIONS
Lab Data
Lab Data Screen
You can factor lab data into the calculation of derived parameters. The monitor
imports laboratory data automatically, through an MIB interface, or after a manual
request through the network.
To display the Lab Data screen
1. Press the
Fast Access
fixed key.
2. Click on
Lab Data
to display most recently captured values.
3. Click on
Capture Labs
to display current values.
MVe
Expired Minute Volume
(TVe x RR) / 1000
L/min
MV/CO
Ventilation Cardiac Output ratio
MV alv / CO
None
O
2
ER
Oxygen Extraction Ratio
(CaO2 - CvO2)/ CaO2
None
P(A-a)DO
2
Alveolar-arterial Oxygen Difference
iO2 x (PB -47) - PaCO2 -
PaO2
mmHg
Qs/ Qt
Intrapulmonary Right-left Shunt
(Percentage shunt)
1/(1 + C(a-v)O2 /
P(A-a)O2 x .003
%
TVd phys
Tidal Volume deadspace
(physiological)
TVe x (1-PeCO2/PaCO2
ml
TVd/ TV
phys
Ratio of Tidal Volume deadspace to
Tidal Volume deadspace
(physiological)
Vd / TVe
not applicable
VO
2
Oxygen Consumption
avDO
2
x CO x 10
mL/min
VO
2
I
Oxygen Consumption Index
VO
2
/ BSA
mL/min/m
2
NOTE:
The displayed parameters vary according to the type of blood analysis device
is connected to the monitor.
Label
Parameter Description
Equations
Units
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...