Modifying Alarm Functions
1. Access the
Alarm Limits
table (see page 5-6).
2. Scroll to the parameter whose alarm functions you wish to configure and
click.
3. Scroll to the alarm function you wish to modify (the first column,
Alarms
, is
highlighted when you first click on the parameter).
4. Choose the new setting, and click to confirm your selection.
5. Repeat steps 2 - 4 for each change.
STVM/STVCM 0 to 45 mm (0 to 4.5 mV)
See “ST Alarms Table” on page 10-10.
TOF-Cnt
0 to 4
Off
Low 0
High 4
TruST Parameters (see: ST TruST
Leads)
Temperature
(Ta/b, T2a/b,
T3a/b)
-5.0 to 50.0
o
C (23.0 to 122.0
o
F)
Off
Low 34.0
o
C (93.2
o
F)
High 39.0
o
C (102.2
o
F)
Δ
T/
Δ
T2/
Δ
T3
-32.0 to 35.0
o
C (-25.6 to 95.0
o
F)
Off
Low 0.0
o
C (32.0
o
F)
High 2.0
o
C (35.6
o
F)
tpCO
2
/pCO
2
2* 10 to 150 mmHg (1.3 to 20.0 kPa)
On
Low 4.0 mmHg (0.5 kPa)
High 6.7 mmHg (0.9 kPa)
tpO
2
/tpO
2
*
10 to 300 mmHg (1.3 to 40.0 kPa)
On
Low 6.7 mmHg (0.9 kPa)
High 10.7 mmHg (1.4 kPa)
PCCO
0.50 to 25.00 L/min
Off
Low 5.50 L/min
High 9.50 L/min
PCCI
0.75 to 15.00 L/min
Off
Low 3.00 L/min
High 8.00 L/min
NOTE:
Turning a parameter’s alarm function on allows alarms to be triggered for that
parameter. It is not a prerequisite for a parameter to be on display or to be
connected for it to produce an alarm.
Turning a parameter’s alarm function off prevents the parameter from
triggering an alarm.
Parameter
Predefined Alarm Range
Default
State
Default Alarm
Setting / Alarm Priority
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...