![Dräger Infinity Delta Instructions For Use Manual Download Page 363](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779363.webp)
H
ARDWARE
S
ETUP
VF8
D
ELTA
/D
ELTA
XL/K
APPA
21-5
Selecting a Sensor
Select a sensor according to your monitoring needs and the size of your patient’s
endotracheal tube. To monitor respiratory mechanics only, use a Flow sensor. For both
respiratory mechanics and etCO
2
parameters, use a combined CO2/Flow sensor.
To monitor respiratory mechanics only
1. Select an appropriate flow sensor.
2. Plug the sensor tubing into the etCO
2
/Respiratory Mechanics pod.
3. Connect the flow sensor to the breathing circuit as shown in the following
illustration. A message at the top left of the screen confirms the size of the
sensor and its correct connection.
Sensor Type
ETT Size
Tidal Volume
Patient Age
Flow
(Neonatal)
< 4 mm
1 - 100 ml
Not specified
Flow
(Adult/Pediatric)
> 4 mm
200 - 3000 ml
Not specified
Combined CO2/Flow
(Neonatal)
2.5 - 4 mm
1 - 100 ml
< 4 years
Combined
CO2/Flow
(Pediatric)
3.5 - 6 mm
30 - 400 ml
2 - 18 years
Combined
CO2/Flow (Adult/
Pediatric)
5.5 mm
200 - 3000 ml
> 12 years
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...