The monitor determines initial and maximum cuff inflation pressure and inflation time
according to the category of the patient, as shown in the following table:
Status Messages
Inflation
Adult
Pediatric
Neonatal
Initial and Maximum
Inflation Pressure (mmHg)
80 + 5
60 + 4
40 + 3
Inflation Time (sec)
120 + 5
120 + 5
60 + 2.5
NOTE:
Perform Venous Stasis on a different arm from that used to measure SpO
2
to
assure proper SpO
2
monitoring.
Message
Possible Cause
Suggested Action
NBP s/d/m > #
NBP s/d/m < #
The NBP value (systolic, diastolic,
mean) exceeds alarm limits.
• Check the patient and treat, if
necessary.
• Change current alarm limits for
patient.
NBP Low
Inflation Limit
The patient’s systolic pressure is higher
than the maximum allowed inflation
limit.
Select the next higher NBP inflation
limit setting.
NBP
Check Cuff Size
The patient’s pulsations are too small to
determine blood pressure.
Check the cuff size.
Move the cuff to another limb.
NBP: Venous
stasis started
Venous stasis mode is enabled.
No action is required.
NBP: Venous
stasis ending
There are 10 seconds remaining in
venous stasis mode.
No action is required.
NBP: Venous
stasis ended
Venous stasis mode is disabled or
completed.
No action is required.
NBP Check
Hose Connection
• The pressure cannot be maintained in
the cuff.
• The inflation time is too short due to a
blocked or kinked hose.
• Check connection between cuff and
hose for debris.
• Check hose and cuff for obstructions
or kinking and replace, if necessary.
NBP Hose
Unplugged
The NBP hose is unplugged.
Reconnect the hose.
NBP Mean Only
The patient's pulse is too low for the
monitor to derive systolic and diastolic
pressure values but large enough to
report a mean pressure value.
• Check the patient and treat, if
necessary.
• Check the hose and cuff.
• Check size and placement of cuff.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...