![Dräger Infinity Delta Instructions For Use Manual Download Page 278](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779278.webp)
16-6
D
ELTA
/D
ELTA
XL/K
APPA
VF8
16 C
ALCULATIONS
The monitor automatically calculates the following hemodynamic parameters:
Label Parameter
Description
Derivation
Units
BSA
Body Surface Area
Boyd or DuBois equation
NOTE:
Boyd equation
:
for patients
whose weight is less than15 kg and
whose height is less than 80 cm:
BSA = WT
(0.7285 - 0.0188 x(log10WT)
x
HT
0.3
x 0.0003207
DuBois equation:
for all other patients:
BSA = WT
0.425
x HT
0.725
x 0.007184
Body weight in grams; height in
centimeters
m
2
CI,
CCI,
ICI
Cardiac Index
(continuous,
intermittent)
CO / BSA
L/min/m
2
p-CI
Thermodilution Cardiac
Index via the PiCCO
pod
p-CO / BSA
L/min/m
2
PCCI
Pulse Contour Cardiac
Index via the PiCCO
pod
PCCO / BSA
L/min/m
2
LHCP
P
Left Heart Coronary
Perfusion Pressure
ART D - PWP
mmHg
LVSW Left Ventricular Stroke
Work
0.0136 x (ART M- PWP) x SV
g x m/beat
LVSWI Left Ventricular Stroke
Work Index
0.0136 x (ART M - PWP) x SVI
g x m/m
2
/beat
PVR
Pulmonary Vascular
Resistance
80 x ((P AM-PWP) /CO)
dynes x sec x cm
-5
PVRI
Pulmonary Vascular
Resistance Index
80 x ((P AM-PWP) / CI)
dynes x sec x cm
-5
x m
2
RPP
Rate Pressure Product ART S x HR
mmHg/min
RVSW Right Ventricular
Stroke Work
0.0136 x (PA M - CVP) x SV
g x m/beat
RVSW
I
Right Ventricular
Stroke Work Index
0.0136 x (PA M - CVP) x SVI
g x m/m
2
/beat
SV
Stroke Volume
CO x 1000 / HR
ml
SVI
Stroke Volume Index
SV / BSA
ml/m
2
SVR
Systemic Vascular
Resistance
80 x (ART M - CVP) / CO
dynes x sec x cm
-5
SVRI
Systemic Vascular
Resistance Index
80 x (ART M - CVP) / C.I.
dynes x sec x cm
-5
x m
2
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...