C
LEANING
, C
ALIBRATION
AND
V
ERIFICATION
VF8
D
ELTA
/D
ELTA
XL/K
APPA
19-15
Sensor Calibration and Verification
For accurate readings, calibrate the Capnostat every time you connect it to a different
etCO
2
module or pod. Calibration is unnecessary if you disconnect the sensor and
later reconnect it to the same device. You can display the date and time of the last
Capnostat sensor calibration by clicking on
Sensor Cal.
in the etCO
2
menu. Verify
the sensor calibration periodically to check that it is functioning correctly, or when
you suspect changes or inaccurate readings.
To calibrate and verify the Capnostat sensor
1. Make sure the monitor is turned on and properly connected to the module or
pod.
2. The monitor displays a message informing you that the sensor is warming up
(~2 minutes at room temperature).
3. When the sensor reaches a stable temperature, the monitor instructs you to
place the sensor on the zero cell.
4. Locate the zero and reference cells on the sensor cable.
5. Place the sensor on the zero cell. The calibration process begins automatically
and takes about 20 seconds.
6. When calibration is complete, the monitor instructs you to place the sensor on
the reference cell.
7. Place the sensor on the Reference cell. The monitor displays the message
etCO2 Verifying Sensor Cal
.
8. When the verification is complete, the monitor displays the message
etCO2
Sensor Cal. Verified.
You can now use the sensor.
If verification fails, the monitor again displays the message e
tCO2 Place Sensor on
Zero Cell
. In this case, repeat the process from step 5. A status message appears if
calibration fails again (see table at the end of this chapter).
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...