M
ONITORING
S
PECIFICATIONS
VF8
D
ELTA
/D
ELTA
XL/K
APPA
B-55
N
2
O:
1, 2
±2 vol % + 8 % rel. (when RR
≤
75 with I:E of 1:2)
O
2
:
1, 2
±3 vol % (when RR
≤
60 with I:E of 1:2)
CO
2
:
1
,
2
±0.5 vol % or ±12 % rel., whichever is greater (when RR
≤
75
with I:E of 1:2)
Drift of measurement accuracy:
≤
1.6 mmHg over 8 hours
Quantitative effects of humidity and
condensate on accuracy
Full accuracy specifications are maintained for all non-
condensing humidity levels.
Adverse effects on performance of
cyclical pressures up to 10 kPa (100
cmH2O)
No effect on accuracy
Gas leakage:
<1.6 ml/min
Nominal rise time:
(10% to 90% at
150 mL/min. flow rate)
CO2:
<350 ms
O2:
<600 ms
N2O:
<500 ms
Agents: <500 ms
Delay time with water trap and 2.5m
sample line:
< 4 sec.
Respiration rate (RRc):
Display: Respiration rate (rpm)
Measurement Range:0 to 90 rpm
Resolution:1 rpm
Accuracy:0 to 60 rpm: ±1 rpm
>60 rpm: not specified
1
Due to the response time of the sensors and the gas sample flow rate, the stated accuracy of O2,
CO2, N2O and anesthetic agents is limited by respiratory rate and inspiratory to expiratory (I:E)
ratio. The effect of respiratory rate and I:E ratio settings on accuracy were determined in a
simulated breathing system using square wave gas concentration waveforms.
2
Auto zero calibrations automatically compensate for drift of measurement accuracy.
3
Gas measurement accuracy is unaffected at the gas levels listed in the table below:
Gas measurement accuracy table (See 3, above)
Gas or Vapor
Gas Level
(Gas levels in % volume fraction)
Nitrous oxide
60
Halothane
4
Enflurane
5
Isoflurane
5
Sevoflurane
5
Xenon
not intended for use with Xenon
Helium
50
Metered dose inhaler
propellants
not intended for use in presence of metered dose inhaler
propellants
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...