9. Return the sensor to the incubator. A status message appears if calibration
fails. Refer to the table at the end of this chapter for status messages.
Status Messages
Message
Condition
Suggested Action
FiO
2
< #
The O
2
value exceeds the upper
alarm limits.
• Check the patient and treat, if
necessary.
• Recalibrate the system.
FiO
2
> #
The O
2
value fell below the lower
alarm limits.
FiO
2
Cal. Canceled
The calibration has been
stopped.
Try to calibrate again.
FiO
2
Cal Failure
The monitor could not calibrate
the FiO
2
sensor.
Recalibrate. If the message persists, try
a new sensor.
FiO
2
Cal Accepted
The calibration was successful.
Informational message only.
FiO
2
21% Cal. in
Progress
The monitor is performing a 21 %
(1-point) calibration.
Wait.
FiO
2
100% Cal. in
Progress
The monitor is performing a
100 % (2-point) calibration.
Wait.
FiO
2
Cal Required
The sensor needs to be
calibrated.
Perform a calibration.
FiO
2
Cal Paused
The monitor is waiting for the
sensor to be exposed to room air
(during 2-point calibration).
Remove the sensor from the T-piece
and expose it to room air.
FiO
2
H/W Failure
Hardware malfunction.
Disconnect the NeoMed pod, then
reconnect it. If the message persists,
return the pod to hospital’s technical
personnel and try a new one.
FiO
2
Pod Unplugged
The monitor cannot detect a
NeoMed pod.
Check the connections and verify the
pod is correctly plugged into the monitor.
FiO
2
Sensor
Unplugged
The monitor is not detecting a
sensor.
Check the sensor connections.
FiO
2
Sensor Failure
The sensor is not accurately
measuring oxygen.
Try the sensor again. If the message
persists, try a new sensor.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...