Calibrating
Mercury calibration is in progress. Complete the calibration before you begin
monitoring the patient.
<xx> Cal.
Accepted
Mercury calibration succeeded
or
the user-entered calibration factor
was accepted.
No action is required.
<xx> Cal.
Failed - Not
Static
The mercury calibration failed
because the input pressure was
not static.
• Make sure the transducer is closed to
patient.
• Check for leaks.
• Keep all tubing motionless.
• Rezero the transducer.
• Refer to the calibration procedures (see
page 14-5).
• Loosen and retighten the transducer dome,
then rezero the transducer.
• Replace the transducer.
<xx> Cal.
Failed - Out of
Range
The mercury calibration failed
because the measured value was
too high or too low.
• Make sure the transducer is zeroed, then
retry. If the retry fails, replace the transducer.
• If calibration requires a factor outside this
range, replace the transducer.
<xx> Zero
before Cal.
During calibration, more than 5
minutes have elapsed since the
last successful zeroing of the
transducer.
Zero the transducer.
<xx> H/W
Failure
IBP channel hardware failure.
• Check the hardware and replace it, if
necessary.
• Call the hospital’s technical personnel or
DrägerService.
Hemo Pod [n]
Disconnected
The hemo pod [1, 2, or 3] is not
connected to the monitor.
Check the cables and connections and
replace it, if necessary.
NOTE
: <xx> represents the IBP parameter label associated with the displayed message.
Message
Possible Cause
Suggested Action
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...