Non-Invasive Blood Pressure (NBP)
Parameter display:
Systolic, Diastolic, Mean
Measurement method:
Oscillometric technique
Modes of operation:
Manual (single measurement), Continuous (5 minutes), or Interval
Interval times:
1, 2, 2.5, 3, 5, 10, 15, 20, 25, 30, 45, 60, 120, and 240 min
Measurement range (heart
rate)
30 to 240 bpm
Pressure measurement
range
(Adult - 270 mmHg):
Systolic NBP: 30 to 250 mmHg
Mean NBP:
20 to 230 mmHg
Diastolic NBP: 10 to 210 mmHg
Pressure measurement
range
(Pediatric - 180 mmHg):
Systolic NBP: 30 to 170 mmHg
Mean NBP:
20 to 150 mmHg
Diastolic NBP: 10 to 130 mmHg
Pressure measurement
range
(Neonatal - 140 mmHg):
Systolic NBP: 30 to 130 mmHg
Mean NBP:
20 to 110 mmHg
Diastolic NBP: 10 to 100 mmHg
Connections:
Quick-release hose connector with single airway
Default inflation pressure:
Adult (270):
160 mmHg ±10mmHg
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Inflation pressure after a
valid measurement
(±10 mmHg):
Adult (270):
Previous NBP
SYS
+ 25 mmHg
Pediatric (180): Previous NBP
SYS
+ 25 mmHg
Neonatal (140): Previous NBP
SYS
+ 30 mmHg
Inflation pressure
after an alarm:
Adult (270):
160 mmHg ±10 mmHg
Pediatric (180): 120 mmHg ±10 mmHg
Neonatal (140): 110 mmHg ±10 mmHg
Maximum inflation
pressure:
Adult (270):
265 mmHg ±5 mmHg
Pediatric (180): 180 mmHg ±10 mmHg
Neonatal (140): 142 mmHg ±10 mmHg
Minimum inflation
pressure:
Adult (270):
110 mmHg ±10 mmHg
Pediatric (180): 90 mmHg ±10 mmHg
Neonatal (140): 70 mmHg ±10 mmHg
Maximum measurement
time:
Adult (270):
2 min ±1 sec
Pediatric (180): 2 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Maximum measurement
time including a retry:
Adult (270):
3 min ±1 sec
Pediatric (180): 3 min ±1 sec
Neonatal (140): 90 sec ±1 sec
Software safety cut-off:
Adult (270):
273 ±3 mmHg
Pediatric (180): 215 ±3 mmHg
Neonatal (140): 153 ±3 mmHg
Hardware safety cut-off:
Adult (270):
300 ±30 mmHg
Pediatric (180): 300 ±30 mmHg
Neonatal (140): 157 ±8 mmHg
Static cuff accuracy:
±3 mmHg
Calibration range:
Adult and Pediatric:10 to 260 mmHg ±3 mmHg
Neonatal:
10 to 150 mmHg ±3 mmHg
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...