Cable Type
When
set
to
Auto Detect
, the
monitor detects the number of
connected lead wires via a
MultiMed pod automatically. If
Auto Detect
mode does not
detect the connected lead wires
(the ECG waveform is not
displayed or the message
ECG
Leads Invalid
or
V+ Lead OFF
appears in the upper left corner
of your screen), you can
manually select the cable type.
The setting
10
(
12 Lead
)
denotes a combination of a
6-lead wire set and a 4-chest
lead wire set, or a 5-lead wire
set and 5-chest lead wire set for
12-lead monitoring.
NOTE
:
Verify the
Cable Type
setting whenever you begin
ECG monitoring (the monitor
remembers the previous
Cable
Type
setting).
NOTE
: The TruST selection
becomes unavailable when you
connect a MultiMed 6, MultiMed
Plus or a MultiMed Plus OR
cable and the
Cable Type
setting is set to either:
3
, or
5.
NOTE
: The selection 10
(12 lead) is only available when
a MultiMed 12 pod is
connected.
•
Auto Detect (default)
– the monitor
compensates automatically for one
disconnected neutral lead. (This feature is
available only with the
Auto Detect
setting.)
• 3
• 5
• 6
• 10 (12 Lead)
NOTE
: The selected cable type determines how
many ECG leads are processed regardless of
which ECG cable is connected. For example, if
you connect a MultiMed 6, MultiMed Plus or
MultiMed Plus OR with a 6 lead wire set, but the
Cable Type
is set to ‘
5
’, the monitor will only
process and display 7 leads. The V+ lead will
not be displayed, trended or broadcast to the
network in this case. However, the V+ lead will
still be displayed in the Show All screen. To
display all 8 leads you must set the
Cable Type
to ‘
6
’.
ECG processing for QRS detection and
arrhythmia are still controlled with the QRS/ARR
Select menu setting (see page 8-23) and the
Lead Setup menu selections for
Channel 1
and
Channel 2
.
The
ECG Options
Submenu
This submenu allows you to configure the following functions.
Filter
Controls the channel bandwidth
and displays a banner in the
ECG1 channel if the selection is
set to
OFF
or
ESU.
No banner is displayed if you
select
Monitor
.
CAUTION
: When the ESU
block or the MultiMed Plus OR
cable is in use, the detection of
pacemaker spikes could be
compromised.
• Click on
OFF
for
maximum bandwidth and
greatest sensitivity to noise or artifact.
• Click on
Monitor
(default) to reduce baseline
drift, muscle artifact, and power line
interference (recommended for standard
monitoring, display, recording, and analog
output).
• Click on
ESU
to reduce signal distortion from
electrosurgical units (see page 8-6 for
information on electrosurgical safety).
Menu
Selection
Description
Available Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...