Electrosurgery
Observe the “Safety Considerations” section on page 6 and the “Electrosurgery”
section on page 11 of these Instructions for Use for general safety precautions during
electrosurgical procedures to reduce ESU interference and enhance user and patient
safety.
The monitor’s ECG function is protected against high-frequency interference from
defibrillators and electrosurgical units.
ESU Block
The MultiMed Plus OR cable and the 5/6-lead ESU block have an integrated electro-
surgical unit (ESU) filter which helps to reduce HF surgery artifacts for improved
reliability. The ESU filter helps to protect the patient from burns caused by ESU-
induced currents flowing through unprotected leads. See chapter 29, for information
on cleaning the device.
NOTE:
Place the electrodes as far from the surgical incision as possible while
maintaining a clinically useful configuration.
Place the cable and lead wires as far from the ESU as possible and
perpendicular to the ESU cables.
Use ESU-return electrode with largest possible contact area. Whenever
possible, place the ESU-return electrode directly under the surgical site,
avoiding bony protrusions.
Replace ECG electrodes regularly.
Read the operating instructions that came with the ESU block for additional
information.
NOTE:
Only use the ESU block during electrosurgery. Only for use with the
MultiMed 3/5, or MultiMed 6.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...