M
ONITORING
S
PECIFICATIONS
VF8
D
ELTA
/D
ELTA
XL/K
APPA
B-57
BIS Parameters
Via BISx Pod
Temperature
BIS index:
Parameter Label:BIS
Measurement Range:0 to 100
Display Resolution:1
Signal quality index:
Parameter Label:SQI
Measurement Range:0 to 100%
Display Resolution:1
Suppression ratio:
Parameter Label:BSR
Measurement Range:0 to 100%
Display Resolution:1%
EMG:
Parameter Label:EMG
Measurement Range:0 to 100 dB (in the frequency range from 70-110
Hz)
Display Resolution:1 dB
Total power:
Parameter Label:PWR
Measurement Range:40 to 100 dB (frequency range 0.5-30 Hz)
Display Resolution:1 dB
Spectral edge
frequency:
Parameter Label:SEF
Measurement Range:0.5 to 30 Hz
Display Resolution:0.1 Hz
Burst Count:
Parameter Label:BCT
Measurement Range:0 to 30
Display Resolution:1
Parameter display:
Absolute temperature, delta temperature (with hemodynamic pods)
Measurement range:
Absolute:–5 to 50 °C (23 to 122 °F)
Delta:
0 to 55 °C (32 to 131 °F)
Resolution:
0.1 °C
Accuracy:
Absolute:±0.1 °C
Delta:
±0.2 °C
Average response time:
<2.5 seconds
Probe accuracy:
±0.1 °C 0 to 50 °C
Degree of protection
against electric shock:
Type CF
Defibrillation protection:
Defibrillation-proof applied part per IEC 60601-1
NOTE:
Range and accuracy values are also applicable to hemodynamic pods.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...