Via Trident Pod
Desflurane
15
Ethanol
3% blood alcohol level
Isopropanol
1
Acetone
0.1
Methane
3
Stimulation current:
Auto mode: read-only; displayed value is based on reference
measurement at startup (max. 60 mA into impedances of 0 to 6.6 k Ohms)
Manual mode: 5 mA (default) to 60 mA, in incr. of 5mA
Pulse width:
100, 200 (default) or 300
μ
s (Monophasic)
Measurement range:
NMT-Temp:20 to 40 °C
Single: 0 to 200 % (of reference twitch)
TOF-Ratio:4 to 150 %
TOF-Count:0 to 4 (number of twitches)
(if TOF-Ratio <4 % or number of twitches <4)
PTC:
0 to 20 (number of twitches)
Measurement
interval:
None, 1s, 10s, 20s (default), 1min, 5min, 15min, 30min
Accuracy:
NMT-Temp:±1 °C
Single: ±10 %
TOF-Ratio:±10 %
Display resolution:
NMT-Temp:1 °C
Single/TOF-Ratio:1%
TOF-Count/PTC:1
Trend scale range:
NMT-Temp:20 to 40° C (incr. of 1 °C)
Single: 0 to 200 % (incr. of 5 %)
TOF-Ratio:0 to 150 % (incr. of 5 %)
TOF-Count:0 to 4 (incr. of 1)
PTC:
0 to 20 (incr. of 1)
Gas measurement accuracy table (See 3, above)
Gas or Vapor
Gas Level
(Gas levels in % volume fraction)
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...