2-14
D
ELTA
/D
ELTA
XL/K
APPA
VF8
2 M
ONITOR
S
ETUP
Unit Manager
The Unit Manager menu lets supervisory personnel configure monitoring functions
for the clinical staff. Access to this menu is restricted by a password. To open the Unit
Manager menu:
1. Press the
Menu
fixed key to open the Main menu.
2. Click on
Monitor Setup
.
3. Click on
Unit Manager
. A data entry box appears.
4. Scroll through the numbers and click successively on the single digits of the
clinical password. If you make a mistake, click on
Backspace
and try again.
5. Click on
Accept
to open the Unit Manager menu. Available functions are
described in the table that follows.
Pulse Tone
Volume
Sets the pulse tone volume
(see page 2-8).
• 10 % - 100 %
The Unit Manager Menu
Menu Item
Description
Available Settings
The Alarm Control Submenu
This menu allows the unit manager to configure alarm annunciation. Open the Unit Manager menu,
click on
Alarm Control
, then follow the procedures outlined in this table to execute the indicated
functions.
All Alarms OFF
Key
Determines whether clinical personnel
can suspend alarms by using the
All
Alarms Off
fixed key on the front of the
monitor.
•
ON (default)
– Pressing
All
Alarms Off
fixed key suspends
alarms.
•
OFF
– Pressing
All Alarms Off
fixed key triggers an error tone.
All Alarms OFF
Time
Sets the alarm suspension time.
WARNING
:
If No Timeout is assigned
to the ‘alarm off period’, no counter
appears and alarms remain
deactivated until you turn them on
again.
•
1, 2 (default), 3, 4,
or
5 min
– A
timer at the top of the screen
indicates the amount of time
remaining in
All Alarms OFF
Time.
•
No Timeout
– Alarms are
suspended indefinitely; no timer
appears.
OR Function
Description
Settings
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...