![Dräger Infinity Delta Instructions For Use Manual Download Page 387](http://html.mh-extra.com/html/drager/infinity-delta/infinity-delta_instructions-for-use-manual_2526779387.webp)
D
ISPLAY
F
EATURES
VF8
D
ELTA
/D
ELTA
XL/K
APPA
22-3
Display Features
FiO
2
values are displayed in a parameter box as shown below.
FiO
2
Setup
The monitor acquires FiO2 signals from the sensor via the NeoMed pod.
To connect the pod to the sensor
1. Set the patient category on the monitor to
Neonatal
.
2. Plug an FiO
2
sensor cable into the FiO
2
connector on the NeoMed pod.
3. Plug the NeoMed pod into the MultiMed
connector on the monitor.
4. Attach a sensor into the FiO
2
sensor cable.
Push the sensor firmly into cable receptacle
until you hear it click.
5. Place the sensor in the incubator or under the
oxygen hood.
1
FiO
2
value and label
2
Alarm limits (crossed bell icon if
alarm is disabled)
3
Attention icon (appears when
the alarm value falls below
21 %)
FiO2 connector
NeoMed Pod
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...