For N
2
O measurements, accuracy is maintained to a respiratory rate of 75
BPM with an I:E ratio of 1:2.
For anesthetic agents, accuracy is maintained to a respiratory rate of 60 BPM
with an I:E ratio of 1:2.
The effect of respiratory rate and I:E ratio settings on accuracy were determined in a
simulated breathing system using square wave gas concentration waveforms.
Precautions
Refer to the general precautions section for etCO
2
/gas analysis.
WARNING:
Clinicians should adjust the fresh gas supply as
needed to correct for the reduction in delivered patient
tidal volume due to sampling of respiratory gases.
Monitoring of oxygen concentration is necessary
during low flow anesthesia due to accumulation of
exhaled metabolic products. If the oxygen
concentration is decreasing, flush the breathing circuit
with fresh gas at regular intervals.
The presence of aerosols in the breathing circuit
should be avoided (displayed agent concentration may
be affected and/or watertrap membrane may be
affected).
Always use Dräger-specified Scio polypropylene
sampling tubing. Never use standard pressure-sensor
tubing (PVC) because it absorbs anesthetic agents,
which are released later (degassing) resulting in
erroneous agent concentration readings.
CAUTION:
To prevent damage to the water trap and measuring system do not use the water
trap with nebulizers; do not allow alcohol, cleaning agents or disinfectants to
enter the water trap; do not wash or sterilize the water trap.
The presence of organic cleaning solutions or gases containing freon will impair
the accuracy of the Scio module.
Mechanical vibrations or shock can adversely affect the accuracy of the Scio
module’s gas measurements.
Under extreme monitoring conditions (and if a network functionality is in use)
intermittent spikes may be present on the Scio waveform. Parameter box data is
not affected.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...