9-6
D
ELTA
/D
ELTA
XL/K
APPA
VF8
9 A
RRHYTHMIA
M
ONITORING
Channel - Lead Selection
Appropriate lead selection is essential for accuracy in arrhythmia monitoring. Ideally,
you should assign the two best leads to the top two channels on the monitor. See page
8-19 for more detailed information.
Processing options are:
ECG1
(
single-channel option)
–
Dedicates processing to the lead that occupies
top channel on the monitor screen.
ECG 1 & 2
(
dual-channel option)
–
Instructs the monitor to determine heart
rate and arrhythmia based on the leads that occupy the two top channels on
the monitor screen.
To configure the monitor for single- or dual-channel monitoring
1. Click on the ECG parameter box to display the ECG menu.
2. Click on
QRS/ARR Select
.
3. Select
ECG1
or
ECG1 & 2
and click.
Arrhythmia Setup Table
When the monitor is operating in Full arrhythmia mode, the ARR setup table allows
you to configure arrhythmia monitoring according to your patient’s needs. The
monitor can detect all events listed in the first column of the table. Using the
remaining columns, you can modify the attributes of each event. Fields that are not
applicable for a certain event category are blank, while those that cannot be modified
are ghosted.
To access the ARR Setup table
Click on the ECG parameter box on the main screen
or
1. Press the
Alarm Limits
fixed key.
2. Click on
ARR
at the bottom right of the Alarm Limits table. The Arrhythmia
Setup menu appears.
NOTE:
The PVC/min current value is only displayed if the monitor is in Full
arrhythmia mode. The PVC/min limit is set in the Alarm Limits table. Refer to chapter 5
for more information on setting alarm limits.
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...