Networked Alarms
The monitor can broadcast networked alarm messages to any compatible monitor or
central station within the network (typically in less than two seconds). However, a
central station will not annunciate or display any alarms for a connected bedside
monitor whose parameter alarms are turned off. In the Infinity network, you can also
group monitors into separate alarm groups to limit the number of messages posted at a
given device (see page 3-14).
Monitors connected to the network automatically relay alarms to the central station. If
the central station cannot acknowledge an alarm within 10 seconds (because of a
network interruption, for instance), the monitor displays the message,
Network Alarm
Error
,
and sounds a tone at maximum volume (100 %). The alarm volume remains set
at 100% until you change it on the Alarm Limits menu (see page 5-6).
Alarm Priorities
The monitor has three alarm priorities: high, medium and low. Previously, Dräger has
referred to these as life-threatening (L-T), serious (SER), and advisory (ADV) alarm
conditions. You can define alarm priorities for arrhythmia and ST parameters only,
using the arrhythmia setup table (see page 9-6) or the ST alarms table (see page 10-
10). Each alarm priority has its own distinctive auditory alarm signal (alarm tone) and
color.
NOTE:
The network distinguishes between unwanted network interruptions (offline
errors) and the deliberate undocking of a Delta/Delta XL bedside device during Pick and
Go transport operations. Removing a monitor from its IDS or Docking Station does not
trigger a network error alarm.
Alarm Priority
Examples
Alarm Color
High
Asystole, ventricular fibrillation
Red
Medium
Neonatal apnea, patient safety conditions during
NBP measurements
Yellow
Low
Technical conditions such as misapplied
transducer or disconnected lead
White
Summary of Contents for Infinity Delta
Page 76: ...2 26 DELTA DELTA XL KAPPA VF8 2 MONITOR SETUP This page intentionally left blank ...
Page 94: ...3 18 DELTA DELTA XL KAPPA VF8 3 NETWORK APPLICATIONS This page intentionally left blank ...
Page 118: ...5 16 DELTA DELTA XL KAPPA VF8 5 ALARMS This page intentionally left blank ...
Page 140: ...7 14 DELTA DELTA XL KAPPA VF8 7 RECORDINGS This page intentionally left blank ...
Page 166: ...8 26 DELTA DELTA XL KAPPA VF8 8 ECG AND HEART RATE This page intentionally left blank ...
Page 176: ...9 10 DELTA DELTA XL KAPPA VF8 9 ARRHYTHMIA MONITORING This page intentionally left blank ...
Page 188: ...10 12 DELTA DELTA XL KAPPA VF8 10 ST MONITORING This page intentionally left blank ...
Page 198: ...11 10 DELTA DELTA XL KAPPA VF8 11 EEG MONITORING This page intentionally left blank ...
Page 218: ...12 20 DELTA DELTA XL KAPPA VF8 12 RESPIRATION This page intentionally left blank ...
Page 256: ...14 22 DELTA DELTA XL KAPPA VF8 14 INVASIVE BLOOD PRESSURE This page intentionally left blank ...
Page 272: ...15 16 DELTA DELTA XL KAPPA VF8 15 CARDIAC OUTPUT C O This page intentionally left blank ...
Page 290: ...16 18 DELTA DELTA XL KAPPA VF8 16 CALCULATIONS This page intentionally left blank ...
Page 372: ...21 14 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS T NEONATAL FLOW SENSOR ...
Page 384: ...21 26 DELTA DELTA XL KAPPA VF8 21 RESPIRATORY MECHANICS This page intentionally left blank ...
Page 526: ...A 6 DELTA DELTA XL KAPPA VF8 A GLOSSARY This page intentionally left blank ...
Page 584: ...B 58 DELTA DELTA XL KAPPA VF8 B TECHNICAL DATA This page intentionally left blank ...
Page 613: ...POD COMMUNICATION VF8 DELTA DELTA XL KAPPA C 29 ...