12
0x02620040
4
Too many tubes not
venting correctly,
sampling for the CC
probe will be stopped
CAU:
CC Sample Probe
Disabled due to Excess
Pressure Errors
Message ID:
0x02620040
Subsystem:
ccScheduler
Multiple tubes had excessive pressure in
thick stopper closed tubes. CC Sample
Probe Disabled.
•
Check cap piercer for broken blade. Refer to
Replace CTS Blade/Wick (1-Blade Thick CTS
Option)
Replace CTS Blade (1-Blade Narrow
to inspect the blade.
•
Perform CC ODC Alignment.
0x0262004B
4
CC Sample Side
Pausing Due to Sample
Rack Subsystem
Stopped
CAU
: CC Sample Side Pausing
due to Sample Rack
Subsystem Stopped
Message ID
: 0x0262004B
Subsystem
: ccSched
Likely cause is an obstruction in the
sample rack handler. Current In-Process
samples will continue until finished. No
more samples will be loaded.
•
When instrument goes to
Stopped
state, press
the DxC STOP button. Wait for the motors to
release power (about 30 seconds, listen for the
sound of the Smart Modules resetting) then
the system.
•
Check for any obstructions in the sample rack
handler and clear any obstruction found.
•
the system.
•
Reload any incomplete racks.
•
Press the RUN button to resume testing.
0x02660001
4
No Diluent Cartridge
CAU:
No Diluent Cartridge
Message ID:
0x02660001
Subsystem:
RASched
Chem Name:
chem id
Diluent cartridge must be loaded to
performed requested tests. Tests
requiring the diluent cartridge will be
incomplete.
•
Load a DIL1 Cartridge. Refer to
.
0x02680001
4
CC Sample Level Sense
Error
Sample ID:
sample id
Rack ID:
rack id
Rack Position:
rack
position
CAU:
CC Sample Level Sense
Error
Message ID:
0x02680001
Subsystem:
ccScheduler
Sample ID:
sample id
Rack ID:
rack id
Position:
rack position
This error occurs under following
conditions:
•
The bottom of the container is
reached before sample is detected
(e.g., empty cup or tube).
•
Insufficient sample volume in
container.
•
Level sense is triggered sooner than
expected for the given container
type.
•
Check sample for adequate volume
•
Check proper rack assignment for type of
container used (reserved vs. non-reserved).
Refer to
Assigning or Reassigning Reserved
•
Check for fluid in collar wash – fluid in collar
wash can trigger false level sense.
•
Check alignment of sample probe in collar wash.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......