0x02620018
2
CC Reagent Delivery
Subsystem Motion
Error, Pausing
CAU:
CC Reagent Delivery
Subsystem Motion Error,
Pausing
Message ID:
0x02620018
Subsystem:
RaSched
This occurs after 2 motion errors on the
Reagent Delivery Subsystem.
The CC side of the system is placed into
Pausing
then
Stopped
state.
•
When the instrument goes to
Stopped
state,
check for possible interference with the motion
of the reagent probes, reagent mixer or reagent
syringe drive.
•
Make sure the reagent drip tray is installed
properly.
•
Make sure the reaction carousel evaporation
cover is installed properly.
•
Check the reagent syringe to make sure the
plunger is seated correctly.
•
Press the DxC STOP button. Wait for the motors
to release power (about 30 seconds, listen for
the sound of the Smart Modules resetting) then
the system.
•
While the system Homes, check for possible
interference issues.
•
If the system Homes successfully, restart the
run and check the reagent probes and mixer for
possible interferences (e.g., improper cable
routing or not in alignment).
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......