A13914AF
9-89
Maintenance
As-Needed/As-Required Maintenance
9
4
Loosen the screw at drain top (1). Gently grasp the ISE drain top and remove the top. Do not lose
the attached metal drip screen (2). (The top may require slight wiggling back and forth to
remove.)
Figure 9.68
5
Carefully set aside the drain top (with tubing #22 attached).
6
Remove the drain tube (1) from the drain.
Figure 9.69
7
Using four to five cotton swabs soaked with deionized water, clean the inside of the lower drain
and the top of drain.