0x04B2001C
4
MC Sample Level Sense
Error
Sample ID:
sample id
Rack:
rack id
Position:
rack position
CAU:
Sample Level Sense Error
Message ID:
0x04B2001C
Sample ID:
sample id
Rack:
rack id
This error occurs under following
conditions:
•
The bottom of the container is
reached before sample is detected
(e.g., empty cup or tube).
•
Insufficient sample volume in
container.
•
Level sense is triggered sooner than
expected for the given container
type.
•
Check sample for adequate volume.
•
Check proper rack assignment for type of
container used (reserved vs. non-reserved).
Refer to
Assigning or Reassigning Reserved
•
Check for fluid in collar wash – fluid in collar
wash can trigger false level sense.
•
Check alignment of sample probe in collar wash.
0x04B2001E
4
MC Sample Probe
Obstruction
Sample ID:
sample id
Rack ID:
rack
Position:
position
This occurs on first
attempt
CAU:
MC Sample Probe
Obstruction
Message ID:
0x04B2001E
Sample ID:
sample id
Rack:
rack
Position:
position
This error occurs when the MC sample
probe ODC (Obstruction Detection &
Correction) sensor indicates that the
probe is obstructed during sample
aspiration.
•
No operator action is necessary.
•
The system will attempt a second aspiration
from this sample.
0x04B2001F
4
Unrecovered MC
Sample Probe
Obstruction
CAU:
Unrecovered MC Sample
Probe Obstruction
Message ID:
0x04B2001F
This error occurs when there was an
Obstruction present at the MC sample
probe. The probe tried to clear the
obstruction by flushing, but the
obstruction could not be removed. The
MC side of the system goes to a
Stopped
state.
•
Flush the MC Sample probe as per the "As
Needed"
Flush the Sample and Reagent Probe
procedure.
•
If flushing probe is not successful, replace MC
Sample probe. Refer to
.
•
If probe is clear and error continues, check for
pinched Sample line tubing.
•
If the error continues contact Beckman
Customer Support.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......