A13914AF
12-4
Troubleshooting Calibration and Result Errors
MC Calibration
Sample/Reference Deviation
•
The Sample and Reference deviation error checks are used as a measure of noise for
electrolytes.
•
Each Sample ADC and each Reference ADC is actually an average of ten readings.
•
The difference between the high and low values (deviation) of the ten electrode readings (taken
milliseconds apart) must be within the maximum deviation.
•
Maximum deviations are as shown in
below:
Span
•
Span is a measure of sensitivity and verifies that two consecutive calibrator level ADC values are
a minimum distance apart.
•
The difference between the average ADC value of Calibrator 1 and the average ADC value of
Calibrator 2 or between the average ADC value of Calibrator 2 and the average ADC value of
Calibrator 3 must exceed the minimum span limit.
•
Select Calibrator Acceptance Limits under Options in the
Rgts/Cal
screen to view current
calibration limits.
Calibration Reports
The Calibration Report can be a useful tool for identifying issues.
below defines the Fields
in the order they are shown for each chemistry on a typical ISE Module Calibration Report.
Table 12.3
Maximum Allowable Sample/Reference Deviations by Chemistry
Chemistry
Maximum Deviations
Sample
Reference
NA
200
200
K
150
150
CL
200
200
CALC
150
150
Table 12.4
ISE Module Calibration Report Field Definitions
Field
Description
Sample Reference
The difference between the sample and reference ADC values. This is used
for the back-to-back, range, and span checks.
For CO
2
, the sample reference value is the ratio between the sample and
reference readings.
Sample
The average of eight ADC readings of the calibrator sample.
For CO
2
, the sample ADC represents the rate of pH change initiated by the
calibrator sample.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......