A13914AF
5-16
Quality Control
QC Chart (Levey-Jennings)
QC Chart (Levey-Jennings)
Introduction
QC Chart shows the results of a control for a specified period (default is current date) in a graphic
form, showing the position of data points relative to the assigned mean and standard deviation. The
results are listed by date and time, most recent results first. QC Chart is available from either the
hard disk or the diskette.
NOTE
Color flagging is based on assigned mean and SD at time of run. If assigned mean and/or SD are
modified, flag color will not change.
QC Chart retrieves only the 4000 most recent results within the date range specified. If more than 4000
results are within the date range, the system does not print the oldest results. You may need to specify
another date range to retireve the additional results.
Accessing QC Charts
1
Select
QC
from the menu bar.
2
From the
QC
screen, select a currently defined control. If necessary, use the scroll bar to view
additional controls.
3
Select
Chart F7
.
4
Type a Start and End date for the QC Chart in the date range fields. The default Start/End date
is the current date. Press
(
Tab
)
to move between data fields.
NOTE
If no data exists for the dates entered, the message "
No Data Available
" appears.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......