A13914AF
10-3
System Status and Commands
Temperatures
10
Cycle Count Description
Temperatures
Introduction
The actual temperature of each of the various components is shown along with:
•
the valid range.
•
an indication of when a parameter is out of limits.
A
yellow
highlight indicates a warning condition and a
red
highlight indicates an error condition.
IMPORTANT
While the system is in the
Stopped
state, temperatures will not be accurate. If the system boots
directly into the
Stopped
state, no temperatures are shown for the systems that caused the
Stop
state. If
the system goes to
Stopped
from
Standby
or
Running
, the temperatures shown reflect the previous state
and will NOT be updated while the system is
Stopped
.
Show Temperature Status
Follow the steps below to monitor instrument components (for example, reaction carousel, cups,
cuvette wash).
1
Select
Status
from the menu bar.
Table 10.2
Counters
Counter
Description
Modular Chem
Each result for each cup is counted separately. Even if ORDAC is triggered, the
counter only increments for the one result.
ISE
A count of the number of samples run through the ISE module. Whether it is a
single chemistry per sample or a sample running all five chemistries, the ISE cycle
count increments only one time per sample processed through the ISE module.
Cartridge Chem
Each cartridge chemistry reagent dispensed is counted.
Blade Count
•
This is the number of caps pierced by the blade(s).
•
For 1-Blade CTS, the system sets this counter to zero when the blade is
replaced. To replace the blade for 1-Blade CTS, refer to
.
CTS Count
The total number of caps pierced by the Cap Piercer Module.
Sample Count
The total number of sample containers that have been scanned by a sample ID
bar code reader.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...