A13914AF
4-10
Reagent Load/Calibration
Reagent Load
Figure 4.9
Reagent Carousel
5
To load reagent:
•
When the screen prompts you to load the cartridge, open the reagent carousel door. Move
the cartridge in front of the bar code reader. A beep will indicate that the bar code is read.
•
Check the screen to verify reagent data is added to the selected position. If the bar code
does not scan, manually type the information.
6
Close the reagent carousel door.
7
If multiple positions are selected to load or unload, repeat Step 4-6.
8
Before processing samples, calibrate and run QC if needed.
Reagent Parameters
Calibration parameters specific for a particular reagent lot are included with certain multipoint,
non-linear reagent cartridges. These parameters are referred to as Reagent Lot Specific Parameters.
The Reagent Lot Specific Parameters, contained in four bar codes, are included on a card in the
Reagent Kit.
These parameters:
•
are loaded as part of a reagent load whenever the parameters are not found in the database.
1.
Cartridge
2.
Top Positions
3.
Bottom Positions
A011871P.EPS
1
2
3
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......