A13914AF
9-15
Maintenance
Weekly Maintenance
9
3
To access the lower portions of the CC reagent probes (under the attached wash collars) remove
the reaction carousel cover, then:
•
Move the reagent probes from the home position to the maintenance indentation (1).
•
Press the probe down into the depression.
•
Wipe the outside portion of the exposed probe.
Figure 9.5
4
To access the lower portions of the MC and CC sample probes (under the attached wash collars)
remove the sample carousel cover. To continue access:
•
Move the sample probes to a clear space (1), between the sample carousel racks and the
sample carousel tub.
•
Press the probe down into the clear area.
•
Wipe the outside portion of the probe.
Figure 9.6
A015937P.EPS
1