Maintenance
Weekly Maintenance
Figure 9.4
NOTE
The wiper tip should be square to the cuvette. The under surface of the wiper should be flush with
the end of the washer probe.
4
Replace the upper section of the cuvette wash assembly and tighten the thumbscrews finger-
tight. Make sure the wash assembly is seated properly on the mounting pins.
5
Verify that all four wash tower probes are centered over a cuvette. Failure to do so may cause a
flood.
6
Reinstall the cuvette wash station cover.
Clean Total Protein and Albumin Cup Modules (DxC 800 Only)
Total Protein and Albumin modular reaction cups and stir bars should be cleaned weekly, or more
frequently if the chemistry does not calibrate, or when chemistry performance indicates.
IMPORTANT
For convenience, the Total Protein and Albumin Cup Maintenance procedure is also included
with the other modular chemistries in Monthly maintenance. Weekly Total Protein and Albumin Cup
Maintenance should be performed at equally distanced intervals.
IMPORTANT
The system must be in
Standby
in order to access the
Maintenance
mode.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......