A13914AF
9-79
Maintenance
As-Needed/As-Required Maintenance
9
Reinstall the CO
2
Measuring Electrode
CO
2
Measuring Electrode Reinstallation
1
Using a lintless tissue, carefully and thoroughly dry the electrode assembly and electrode port.
Electrode and electrode port must be completely dry before reinstallation of the electrode.
2
Insert the CO
2
electrode onto the electrode port being careful to align key pin on electrode with
keyway in the electrode port. Insert retainer nut and turn until finger-tight.
3
To test for proper seating of electrode, gently pull on electrode body. The electrode assembly
should not move. If the electrode moves, remove it and try to install it again. If installation is
difficult, check for an extra or missing quad-ring.
4
Reconnect tubing lines #33 and #32 to the electrode. Remove hemostats. Make sure the tubings
fit tightly on the electrode port.
5
Reconnect the electrode cable to the appropriate connector.
6
When the necessary electrode service is completed and the electrode is properly replaced,
select
OK
to exit the Maintenance pop-up confirmation window.
NOTE
While the system is priming, observe flow cell for leaks. Stop priming if any leaks are observed.
7
Select
Close
in the
Result
pop-up window to exit the
ISE Service
screen.
8
Adjust the alkaline buffer damper volume. Refer to
Adjust the Fluid Level of the Alkaline Buffer
under Monthly Maintenance, in this chapter.
CAUTION
Pinch hazard. Keep fingers clear of the ISE module as you lower it.
9
Lower the ISE module by releasing the locking pin. Be careful NOT to pinch tubing as the module
is lowered.
10
Reinstall the ISE module cover.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......