A13914AF
12-9
Troubleshooting Calibration and Result Errors
Linear Calibration
12
Calibration Error Log
The fields listed below categorize information about a failed calibration.
Multiple errors are shown here for any chemistry failure. The errors should
correlate with any of the asterisked (*) fields listed within the calibration report.
This section of the report appears only when a calibration fails.
• Reagent
– Identifies the failed chemistry.
• Lot Number
– Identifies the lot number of the reagent which failed
calibration.
• Serial Number (S/N)
– Identifies the serial number of the specific cartridge
which failed.
• Cuvette
– Identifies the cuvette in which the specific replicate failed.
• Condition
– In the event that an error occurs, an asterisk (*) appears next
to the value which is out of specification. The corresponding error flag will
print in this field. Only those errors that FAIL the calibration will print.
• Value
– Shows the absorbance value that failed calibrator or reaction limits.
• Magnitude
– This is an indication of the severity of the failure. Depending
on the error, the magnitude may reflect the actual value of the error or it
may show the absolute difference between the limit and the actual value.
Likewise, the magnitude may not be applicable to the error which has
occurred. If this is the case, no value will appear under this field.
Error Flag
Magnitude
BLANK RATE
absolute delta
BLANK ABS
absolute delta
RXN RATE
absolute delta
RXN ABS
absolute delta
INIT ABS HI
actual value
INIT RATE HI
actual value
BACK TO BACK
absolute delta
OCR
absolute delta
All others
no magnitude
Table 12.6
CC Linear Calibration Report Field Descriptions
(Continued)
Field
Description
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......