A13
9
14
AF
12
-1
3
T
roubl
es
hoo
tin
g
C
a
libr
a
ti
on and
Res
u
lt
E
rr
o
rs
Co
m
m
o
n
Er
ro
r Me
ss
a
g
e
s a
n
d C
o
rr
e
ct
ive
A
ct
io
n
s
12
0x0262001E
2
CC Reagent Probe B
Motion Error
CAU:
CC Reagent Probe B
Motion Error
Message ID:
0x0262001E
Subsystem:
RASched
Action Name:
action name
This error is caused by a Vertical or
Rotary motion error. This will cause the
instrument to go into a
Stopped
state.
•
Raise canopy and check for obvious obstruction
of the CC reagent Probe B.
•
the system.
While system is homing, observe motion of the
CC reagent probe B. The probe should complete
its operations in a smooth manner.
•
If the error continues, check the event log to
determine the specific subsystem error. Refer to
•
If event log indicates a vertical error, check that
the reagent cartridges on the carousel are
pushed in and seated properly.
•
Perform the appropriate vertical or horizontal
CC reagent Probe B alignment. Refer to
.
•
Contact Beckman Coulter Support Center for
additional information.
0x02620020
2
CC Reagent Syringe
Motion Error
CAU:
CC Reagent Syringe
Motion Error
Message ID:
0x02620020
Subsystem:
RASched
Action Name:
action name
This error occurs when there is an error
within the CC Reagent Syringe
Subsystem. This includes both the Shear
valve Motor and the Lead Screw Motor.
•
Raise canopy and check for proper seating of
syringe assembly.
•
the system.
While system is homing, observe motion of the
CC reagent syringe. The syringe should
complete its operations in a smooth manner.
•
If the error continues, contact Beckman Coulter
Support Center for additional information.
0x02620021
2
CC Reagent Switch
Motion Error
CAU:
CC Reagent Switch
Motion Error
Message ID:
0x02620021
Subsystem:
RASched
Action:
This error will occur whenever there is a
motion error in the Reagent Switch (A/B
valve). This error does not occur during a
home, only runtime, primes, etc.
•
the system.
•
Contact Beckman Coulter Support Center for
additional information.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......