A13914AF
10-8
System Status and Commands
Instrument Commands
Instrument Commands
Instrument commands are high level system commands that are used to control the instrument and
printer.
Commands
Select
Instr Cmd
from the menu bar.
The following commands are available:
•
Home
– moves all mechanical assemblies to a known position.
•
Pause
– stops additional tests from starting; any tests with reagent already added will complete.
•
Stop Print
– prevents print requests in the queue from being sent to the printer.
•
Shutdown
– reboots the system (warm boot) or safely powers off the system.
•
Pause Waste B
– allows the external Waste B container to be emptied while the system is
running.
•
Resume Waste B
– resumes the dumping of Waste B to the external container.
•
Enable/Disable Modules
– enables or disables modules.
•
Unload all racks
– unloads all racks.
Home
Introduction
Home is used to move mechanical assemblies to a known "home" position and primes the system.
Why Use Home
Home is useful in the following situations:
•
to attempt recovery from a motion error.
•
to return the system or one component of the instrument from a
Stopped
state to
Standby.
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......