A13
9
14
AF
12
-4
Tro
u
blesho
o
ting Cal
ibr
a
tio
n a
n
d Result
E
rro
rs
Comm
on Er
ro
r Me
ss
ages
and
C
o
rr
ec
tiv
e
Act
ion
s
0x01FA0009
4
DI Water Canister Is Not
Filling At All
CAU:
DI Water Canister Is Not
Filling At All
Message ID:
0x01FA0009
Whenever the DI water canister does not
fill within its given amount of time, it will
time out and flag that it has not filled.
•
Make sure DI water is turned on at the source
and present in the DI Water Reservoir.
•
If Reservoir is full,
—
Press the STOP button and
the
system.
—
If the error continues, contact Beckman
Coulter Support Center for additional
information.
•
If Reservoir is empty,
—
Make sure the Manual water inlet valve is
turned on.
—
Make sure the instrument has not rolled on
top of the water inlet lines.
—
and restart the system.
—
If the error continues, contact Beckman
Coulter Support Center for additional
information.
0x01FA000B
4
Wash Concentrate
Reservoir Is Not Filling
At All
CAU:
Wash Concentrate
Reservoir Is Not Filling At All
Message ID:
0x01FA000B
Wash Concentrate Reservoir Is Not Filling
At All.
•
Load a new bottle of Wash Concentrate. Refer to
Load/Unload Modular Chemistries (MC)
.
0x01FA0014
4
Both The Waste B Exit
Sump And The
Collection Bottle Are
Full. Switch #29,15,9.
CAU:
Both The Waste B Sump
And The Collection Bottle Are
Full
Message ID:
0x01FA0014
Both The Waste B Sump And The
Collection Bottle Are Full. Only applies to
systems configured with the Split Waste
Option.
•
Empty the external waste B collection bottle and
Resume Waste B.
0x01FA0015
4
External Waste B
Collection Bottle Is Full.
Switch #29. Waste B
Paused. Replace Waste
B Container and select
Resume in Instrument
Commands when
ready.
CAU:
External Waste B Bottle Is
Full
Message ID:
0x01FA0015
The Waste B Container is Full. Only
applies to systems configured with the
Split Waste Option.
•
Empty the external waste B collection bottle and
Resume Waste B.
Table 12.10
Error Messages and Corrective Actions
(Continued)
ID
Class #
Pop-up
Event Log
Description
Corrective Action
Summary of Contents for UniCel DxC 600
Page 4: ...A13914AF iv Revision History ...
Page 26: ...A13914AF xxvi Safety Notice Documentation Symbols ...
Page 36: ...A13914AF xxxvi Introduction How to Use this Manual ...
Page 132: ...A13914AF 3 34 System Setup Options Auto Generation of Control ...
Page 190: ...A13914AF 5 26 Quality Control Review Archived Data ...
Page 364: ...A13914AF 9 128 Maintenance As Needed As Required Maintenance ...
Page 384: ...A13914AF 10 20 System Status and Commands Unload All ...
Page 407: ......
Page 490: ...A13914AF Glossary 10 Glossary ...
Page 497: ......