System Description
Cartridge Chemistry: Calibration Theory
Blanked Rate
Chemistries
FOR HIGH CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate
rep1
(Reaction Rate - Blank Rate) = Delta Rate
rep2
(Delta Rate
rep1
+ Delta Rate
rep2
) × 0.5 = Delta Rate
avg
(hi)
FOR LOW CALIBRATOR LEVEL:
(Reaction Rate - Blank Rate) = Delta Rate
rep1
(Reaction Rate - Blank Rate) = Delta Rate
rep2
(Delta Rate
rep1
+ Delta Rate
rep2
) × 0.5 = Delta Rate
avg
(lo)
Offset = Cal Set Point (hi) - [Cal Factor × Delta Rate
avg
(hi)]
Sample values are calculated by the following equation:
[(Reaction Rate - Blank Rate) × Cal Factor] + offset = sample value
IMPORTANT
For single point linear calibration, the low calibrator is a fixed zero point
and the offset is equal to zero.
Table 1.17
Calculation of Calibration Factors for Endpoint and Rate Chemistries
(Continued)