A13914AF
9-59
Maintenance
Six-Month Maintenance
9
IMPORTANT
A new glucose sensor may drift over the first 24 to 48 hours. Always verify control recoveries
after calibration. More Frequent calibration may be required during the first 24 to 48 hours following
installation.
CAUTION
Do not touch membrane tip.
1
Remove the AccuSense glucose oxygen sensor (PN 467432) from the package.
2
Remove the protective cap from the electrode connector.
3
Check the sensor for the presence of the gasket on the membrane tip. If the gasket is missing,
replace it at this time.
4
Visually inspect the sensor for leakage or puncture of membrane.
5
Hold the sensor in a vertical position with the membrane tip pointed down. Tap on the side of
the sensor body to remove any bubbles from the membrane tip.
6
Remove the ISE module and MC module covers.
7
Push the glucose sensor connector into the empty connector port (1) of the BUNm/UREAm
module until the latch clicks.
Figure 9.40